Recall

Daavlin Distributing Company: Photera 400, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders Recall

Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000). Product details: Photera 400, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders

Active Official source

Verification

Official-source record

This page is included in the public index because it is tied to a listed official source. ClaimWatch does not publish trend-only candidates to public SEO pages.

Agency
FDA
Record ID
Z-2742-2026
Canonical
https://claimwatch-mvp.vercel.app/item/daavlin-distributing-company-photera-400-phototherapy-device-ultraviolet-lamp-for-dermatol/

What happened?

Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000). Product details: Photera 400, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders

Who may be affected?

US Nationwide distribution.

What should consumers do?

Affected quantity: 2 units

Matched search terms

No search trend terms attached yet.

FAQ

Is this page official?

No. ClaimWatch links to official sources and marks whether a page is verified or still being monitored.

Is this legal, medical, or financial advice?

No. This page is informational. Always confirm eligibility, deadlines, and safety instructions through official sources.

Similar alerts